歡迎收看高準"咋整"系列視頻
Welcome to the Micro Motion How-To video series
在這一集裡面,我們將介紹如何正確的對高準科氏力流量計進行調零
In this video we will cover how to properly zero a Micro Motion coriolis meter
和"零點驗證"
and introduce Zero Verification
學習的目標如下:
The learning objectives are the following:
理解零點、出廠零點、現場零點和零點驗證
Understand the definitions for zero, factory zero, field zero, and Zero Verification
理解零點驗證功能和如何將其用於您的工藝
Understand the Zero Verification feature and how it is applied to your process
了解如何使用5700變送器表頭和ProLink
Know how to properly zero a coriolis meter
對科氏力流量計進行正確調零
using the 5700 transmitter display and ProLink
完整的指導書
For the complete set of instructions
請參考高準變送器組態和使用手冊
please refer to the Micro Motion transmitter configuration and use manual
第一步:理解零點的定義
Step One: Understanding the Definition of Zero
術語"零點"用於描述流量測量過程的許多部分
The term Zero is used to describe many pieces of the flow measurement process
我們將關注零點標定和零點偏移的細節
We will focus on the specifics of zero calibration and the zero offset
零點的定義是所需要的偏移量
The definition for zero is the offset required
以便在零流量條件下使左右微動傳感器同步
to synchronize the left and right pick off under conditions of zero flow
在實驗室條件下,通過認證的高準標定設備上
The zero established at the certified calibration facilities of Micro Motion
建立的零點稱為出廠零點
under laboratory conditions is known as the Factory Zero
現場零點的定義是
The definition for Field Zero is a zero that is established
在高準標定設施以外的
outside of the calibration facilities of Micro Motion
比如那些在客戶現場安裝後的建立的零點
such as those established after installation at a customer site
零點驗證是一個用於評估已有出廠零點
Zero Verification is a procedure used to evaluate the stored factory zero
並決定是否需要現場零點標定來改善測量精確性的過程
and determine whether or not a field zero calibration can improve measurement accuracy
第二步:驗證零點
Step Two: Verify the Zero
在大多數情況下,工廠調零要比現場調零更精確
In most cases the factory zero is more accurate than the field zero
除非滿足以下條件,否則不要進行流量計的調零:
Do not zero the flow meter unless one of the following is true:
現場流程要求調零
The zero is required by site procedures
已有出廠零點值無法通過零點驗證步驟
The stored factory zero value fails the zero verification procedure
有高嚴重度等級報警的情況下不要驗證零點或進行流量計調零
Do not verify the zero or zero the flow meter if a high severity alarm is active
排除報警原因後驗證零點或進行流量計調零
Correct the cause of the alert then verify the zero or zero the flow meter
在低嚴重度等級報警時,你可以驗證零點或進行流量計調零
You may verify the zero or zero the flow meter if a low severity alarm is active
流量計的準備
Prepare the flow meter
流量計電子部件上電後
Allow the flow meter electronics to warm up for
預熱至少20分鐘
at least 20 minutes after applying power
將工藝流體流過傳感器
Run the process fluid through the sensor
直到傳感器溫度達到正常工藝操作溫度
until the sensor temperature reaches the normal process operating temperature
關閉下遊隔離閥以停止流動
Stop flow through the sensor by shutting a downstream isolation valve
如果可以則關閉上遊隔離閥
and then an upstream isolation valve if available
確認傳感器兩端已阻斷,介質不再流動
Verify that the sensor is blocked in that flow has stopped
傳感器已完全充滿過程流體
and that the sensor is completely full of process fluid
開始零點驗證,等待完成
Start the Zero Verification procedure and wait until it completes
如果零點驗證失敗
If the Zero Verification procedure fails
確認傳感器兩端已完全阻斷,介質不再流動
confirm that the sensor is completely blocked in that flow has stopped
傳感器已完全充滿過程流體
and that the sensor is completely full of process fluid
確認過程流體沒有汽化或冷凝
Verify that the process fluid is not flashing or condensing
且不含可沉澱顆粒物
and that it does not contain particles that can settle out
重複零點驗證流程
Repeat the Zero Verification procedure
如果再次失敗,進行流量計調零
If it fails again, zero the flow meter
第三步:進行科氏力流量計調零
Step Three: Zero the Coriolis Meter
進行零點標定的初始調零條件
The initial zero conditions for a Zero Calibration
與零點驗證時所做的一樣
are the same in which the Zero Verification was performed
如果條件有變化
If your conditions have changed
重新建立初始條件:
Re-establish the initial conditions:
流量計上下遊都用閥隔斷
The meter is blocked in using upstream and downstream valves
確認沒有流量
Confirm flow has stopped
檢查驅動增益、溫度和密度值讀數穩定
Check that drive gain, temperature and density readings are stable
開始調零步驟,等待完成
Start the Zero procedure and wait until it completes
當標定完成,如果調零步驟成功
When the calibration is complete, if the Zero procedure was successful
會顯示標定成功的信息和一個新的零點值
a calibration success message and a new zero value are displayed
如果零點標定失敗,會顯示標定失敗信息
If the zero procedure failed, a calibration failed message is displayed
一旦零點標定步驟成功完成
Once the zero calibration procedure has successfully completed
打開閥門恢復流過傳感器的正常流量
restore normal flow through the sensor by opening the valves
如果調零失敗,確保傳感器中沒有流量然後重試
If the zero fails, ensure that there is no flow through the sensor then retry
移除或減少機電噪聲源然後重試
Remove or reduce sources of electromechanical noise then retry
使用ProLink來驗證零點及進行科氏力流量計調零
Using ProLink to Verify Zero and Zero the Coriolis Meter
在ProLink中,選擇設備工具->標定->智能零點驗證和標定
In ProLink, select Device Tools, Calibration, Smart Zero Verification and Calibration
在這個菜單下,點擊"驗證零點"
In this menu select the Verify Zero button
零點驗證完成後
After the Zero Verification is complete
閱讀驗證的結果和下一步建議措施
read the result of the Verification and the next course of Recommended Action
建議的措施有"建議進行新的零點標定"
The Recommended Actions are a new zero is recommended
或是現有零點無需調整
or the existing zero does not need to be changed
顯示一條"無需新零點"的建議
with a recommendation of No new zero needed
如果建議進行新的調零,點擊"標定零點"
If a new zero is recommended, select the Calibrate Zero button
這會進入"零點標定"菜單
This will open the Zero Calibration menu
在菜單內點擊"標定零點"
Select the Calibrate Zero button in this menu
一旦標定完成
Once the calibration is complete
會顯示一條標定是否成功的信息
it will display a message if the calibration was successful
到這裡,"咋整"視頻的
This concludes the How-To video of
科氏力流量計調零和零點驗證就結束了
Zeroing the coriolis meter and Zero Verification
總結一下,您現在應當:
In summary you should now know the following
理解零點、出廠零點、現場零點和零點驗證的定義
Understand the definitions for zero, factory zero, field zero and Zero Verification
理解零點驗證功能和如何將其用於您的工藝
Understand the Zero Verification feature and how it is applied to your process
知道如何使用5700變送器表頭和ProLink
Know how to properly zero a coriolis meter
對科氏力流量計進行正確調零
using the 5700 transmitter display and a ProLink
完整的指導書
For the complete set of instructions
請參考高準變送器組態和使用手冊
please refer to the Micro Motion transmitter configuration and use manual
在艾默生官網上有哦
which can be found on www.emerson.com